A 10-port valve with integrated pre-column for sample enrichment

An ALEXYS system is used in combination with a solid phase sample pre-concentration step. A 1 mL sample volume is loaded onto a pre-column by a user program in the AS 100 autosampler. For this purpose the AS 100 has been equipped with a 10-port valve. A secondary pump is used to supply the washing solvent.

 

User program for BPA analysis: A 1 mL loop is filled with sample (red). The 10-port valve switches and pump 2 with solvent B (5% ACN, black) carries the sample onto the pre-column. The valve switches back and the pump 1 with solvent A (40% ACN, blue) flushes the concentrated sample from the pre-column onto the analytical column. The blue vial contains washing liquid. Already during analysis, filling of the 1 mL sample loop for the next run is started. This user program is used in 'Bisphenol A'.

Application notes

  • 216-005 - Bisphenol A in drinking water

ALEXYS Bisphenol A

The system has an AS 100 autosampler with 10-port valve for sample enrichment.

 