ALEXYS Bisphenol A

Antec Leyden has developed an integrated system solution for the analysis of Bisphenol A in drinking water or beverages from polymer bottles. The system uses a 10-port valve for automated sample enrichment to improve detection limits.

Bisphenol A

  • Automated sample enrichment for better detection limits
  • Excellent reproducibility and linearity
     

Fig. 1. Analysis of water samples (3 upper traces) and standards (2 lower traces: 0.5 and 1 nmol/L). Only water sample no 1 contained BPA (1.5 nmol/L).

Table 1. System Solution for Bisphenol A in drinking water

Part no.

Description
180.0090A ALEXYS Bisphenol A

Configuration

For the analysis of Bisphenol A an AS 100 cool with 1 mL injection loop and a 10-port injection valve was used. The HPLC system uses an additional pump for sample enrichment on a pre-column. A 2mm column has been used at a flow rate of 200 uL/min which results in less chemical waste. A flowcell with glassy carbon electrode and HyREF was used for analysis. The HyREF is completely maintenance free, and compatible with high modifier percentages. The ALEXYS system comprises the following parts:

  • AS 100 autosampler with 10port valve, cool
  • LC 110 HPLC pump (2)
  • OR 110 organiser rack
  • Clarity data acquisition software
  • DECADE II electrochemical detector
  • Flow cell with GC WE, HyREF

Application notes and posters

  • 216-005 - Bisphenol A in drinking water

ALEXYS Bisphenol A

System with automated sample enrichment using an AS 100 with a pre-column and a 10-port valve.
Read more about the integrated pre-column for enrichment.

 

    System solutions from Antec Leyden .... naturally!