ALEXYS Disulfides Analyser

Antec Leyden has developed integrated system solutions for the analysis of (di-) sulfides such as (di-)glutathione, cystine/cysteine, homocystine/homocysteine. An ALEXYS system has been used with a DECADE II with dual cell control and a reactor cell for reduction of disulfides to thiols. In the second cell thiols are detected by oxidative amperometric detection.

R’-S-S-R” + 2 H+ + 2 e-   -->  R’SH + R”SH

Thiols as well as the oxidised free disulfides can be quantified this way, which is not possible when a pre-column reduction of disulfides is applied.

Disulfides

  • Both mono- and disulfides in one run
  • Detection limit 20 fmol on column (10 nM/2µL)

Fig. 1. Analysis of 1 µM cystine (CSSC), cysteine (CSH), glutathione (GSH), homocysteine (HCSH) and di-glutathione (GSSG) standards (2 µL injection). Disulfides CSSC and GSSG are not visible when switching off the pre-reduction cell (blue trace).

Table 1. System Solutions for Disulfides

Part no.

Description
180.0068A ALEXYS Disulfides 1 for RED-OX of (di-)sulfides

Configuration

For the analysis of disulfides we selected the Flexcell electrochemical cell. This cell has an exchangeable working electrode which is particularly useful in this application. The electrode material tends to vanish over the years, in the process of amperometric detection of thiols. Exchanging an electrode disc only takes a few minutes. The HyREF is completely maintenance free! The ALEXYS system comprises the following parts:

  • AS 100 autosampler, cool
  • LC 110 HPLC pump
  • OR 110 organiser rack
  • Clarity data acquisition software
  • DECADE II electrochemical detector with DCC
  • Flexcell with Au WE, HyREF
  • Reactor with GC WE, HyREF

ALEXYS Disulfides II for homocysteine uses a standard HPLC system, and a Flexcell with gold working electrode.

Table 2. Column for Disulfides

Part no.

Description
250.1114 ALF-115, 150x1.0 mm ID, 3 µm, C18 (disulfides)
250.1122 ALF-515, 150x4.6mm, 3um, C18 (Homocysteine)

Application notes and posters

  • 211-004 - Analysis of glutathione and other (di-)sulfides
  • 211-003 - Homocysteine in plasma

ALEXYS Disulfides 1

The system uses 2 Flexcells in series for reduction of disulfides before oxidative detection of monomers in the second cell.

 