DECADE II Operational Qualification describes the OQ procedure
as advised by the manufacturer. It is a result from our interpretation
of many regulations and laboratory practises. In addition, feedback
from users and representatives helped us to finalize this procedure.
A complete OQ for DECADE II consists of:
- Performance test of DECADE II electronics
- Performance test of DECADE II with flow cell
The second check is an overall HPLC-ECD check running a test sample.
All qualification checks have to be approved, or should be marked
"n/a" if not applicable. Any deviation observed must be documented
in the 'non-conformance' record.
An OQ script in DECADE II Dialogue guides you through all steps
of the procedure. All test results are electronically stored. After
finalising a 4 pages OQ report is generated.
The OQ document describes:
- OQ for DECADE II electronics
Ecell
Output at 10 V / 1 V
Zero IE values
Linearity
Temperature
Dummy cell test
Noise and drift
- OQ for DECADE II with flow cell
OQ Testing procedure
QC criteria
Column properties
- OQ certification
- Non-conformance record
|