DECADE II Operational Qualification describes the OQ procedure
as advised by the manufacturer. It is a result from interpretation
of many regulations and laboratory practises. In addition, feedback
from users and representatives helped to finalize this procedure.
A complete OQ for DECADE II consists of:
- Performance test of DECADE II electronics
- Performance test of DECADE II with flow cell
The second check is an overall HPLC-ECD check running a test sample.
All qualification checks have to be approved, or should be marked
"n/a" if not applicable. Any deviation observed must be documented
in the 'non-conformance' record.
An OQ script in DECADE II Dialogue guides you through all steps
of the procedure. All test results are electronically stored. After
finalising a 4 pages OQ report is generated.
The OQ document describes:
- OQ for DECADE II electronics
Ecell
Output at 10 V / 1 V
Zero IE values
Linearity
Temperature
Dummy cell test
Noise and drift
- OQ for DECADE II with flow cell
OQ Testing procedure
QC criteria
Column properties
- OQ certification
- Non-conformance record
The DECADE II OQ kit (p.n. 250.3000) consists of:
|
Part no.
|
Description |
| 171.0022 |
DECADE II OQ manual |
| 171.9002 |
DECADE II Dialogue, OQ version |
| 250.0150 |
Ecell test cable D conn. |
| 250.0152 |
Output cable BNC |
| 250.0154 |
1.00 MOhm, 0.05% dummy cell D
conn. |
| 250.0156 |
I/O test plug (control board) |
| 250.0158 |
I/O test plug (sensor board) |
| 250.1050 |
HPLC column for OQ |
| 250.1060 |
MOPEG solution for OQ |
| 250.1064 |
Concentrated buffer for OQ |
Return to qualification procedures.
Or enter our download section
for a demo version of DECADE II Dialogue or ALEXYS data system.
The software has limited functionality but can be activated with a
USB license dongle. No further downloads are necessary.
|