ALEXYS Performance Qualification describes the PQ procedure
as advised by the manufacturer. It is a result from interpretation
of many regulations and laboratory practises. In addition, feedback
from users and representatives helped to finalize this procedure.
A complete PQ for ALEXYS consists of:
- Reproducibility test
- Linearity test
- Detection performance test
The 'ALEXYS PQ kit, GC WE'
(p.n. 250.3020) has the following parts:
|
Part no.
|
Description |
| 180.0026 |
ALEXYS PQ manual |
| 250.1050 |
HPLC column |
| 250.1060 |
MOPEG 0.5 µmole/L, 2 mL (3x) |
| 250.1064 |
Concentrated buffer |
| 250.0230 |
AS 100 outlet assembly |
| 250.0232 |
DECADE II inlet assembly |
For systems with a flow cell with gold working electrode a
special PQ kit is available. The 'ALEXYS PQ kit,
Au WE' (p.n. 250.3030) has
the following parts:
|
Part no. |
Description |
| 180.0027 |
ALEXYS PQ manual - for Au WE |
| 250.1050 |
HPLC column |
| 250.1060 |
MOPEG 2 µmole/L, 2 mL (3x) |
| 250.1064 |
Concentrated buffer |
| 250.0230 |
AS 100 outlet assembly |
| 250.0232 |
DECADE II inlet assembly |
Return to qualification procedures.
|